簡要描述:塑料介電擊穿/薄膜耐壓測試儀試驗在試驗箱中進(jìn)行,試驗箱門打開時電源加不到高壓變壓器輸入端,即高壓側(cè)無電壓。100KV測試設(shè)備高壓電極距離試驗箱壁的Z近距離大于270mm,50KV測試設(shè)備高壓電極距離試驗箱壁的Z近距離大于250mm,試驗時即使人接觸箱壁也不會有危險。
產(chǎn)品分類
Product Category詳細(xì)介紹
品牌 | AIRTIMES/中航時代 |
---|
ZJC-100KV計算機(jī)控制塑料介電擊穿/薄膜耐壓測試儀
ZJC-100KV computer controlled voltage breakdown tester
一、概述:
I. overview:
ZJC-100KV塑料介電擊穿/薄膜耐壓測試儀電壓擊穿試驗儀由電腦控制,可實時繪制試驗曲線,顯示擊穿電壓測試數(shù)據(jù),判斷準(zhǔn)確,并可保存,分析,打印試驗數(shù)據(jù),軟件系統(tǒng)操作方便,介電強(qiáng)度設(shè)備擊穿性能穩(wěn)定,安全可靠。
ZJC-100KV voltage breakdown tester is controlled by computer, can draw test curve in real time, test data of breakdown voltage, judge accuray, and can save, analyze and print test data. The software system is easy to operate, and dielectric strength equipment is stable, safe and reliable.
二、符合標(biāo)準(zhǔn):
Two, accord with standard:
1、GB1408.1-2006《絕緣材料電氣強(qiáng)度試驗方法》
1, GB1408.1-2006 "electrical strength test method for insulation materials"
2、GB1408.2-2006《絕緣材料電氣強(qiáng)度試驗方法 第2部分:對應(yīng)用直流電壓試驗的附加要求》
2, "GB1408.2-2006 electrical strength of insulating materials test methods part second: Additional requirements for the application of DC voltage test"
3、GB/T1695-2005《硫化橡膠工頻擊穿電壓強(qiáng)度和耐電壓的測定方法》
3, GB/T1695-2005 "vulcanized rubber power frequency breakdown voltage strength and resistance to voltage measurement method"
4、ASTM D149《固體電絕緣材料工業(yè)電源頻率下的介電擊穿電壓和介電強(qiáng)度的試驗方法》
4, D149 ASTM "test method for dielectric breakdown voltage and dielectric strength of the dielectric breakdown voltage at the industrial power supply frequency of the solid electrical insulation material"
三、主要技術(shù)要求:
Three, the main technical requirements:
1、設(shè)備輸入電壓: AC220
1, equipment input voltage: AC220
2、試驗電壓方式: 交流 0--100 KV ; 直流 0--100 KV;
2, the test voltage: AC 0--100 DC 0--100 KV KV;
3、電器容量:10KVA;
3, electrical capacity: 10KVA;
4、試驗方法:0-100KV全量程可調(diào);
4, test method: 0-100KV full range can be adjusted;
5、擊穿及耐壓試驗升壓速率: 0.1 KV/S 0.2 KV/S 0.5 KV/S1.5 KV/S2 KV/S2.5 KV/S 3.0 KV/S ;
5, the breakdown voltage test and pressure rate: 0.1 KV/S 0.2 KV/S 0.5 KV/S1.5 KV/S2 KV/S2.5 3 KV/S KV/S;
6、試驗方式:直流/交流試驗:1、勻速升壓 2、階梯升壓 3、耐壓試驗;
6, test mode: AC / DC test: 1, 2, speed boost, 3 step-up voltage withstand test;
四、試驗方式:
Four, test method:
1、絕緣試樣空氣中擊穿或耐壓試驗
1, insulation test of air in the air breakdown or pressure test
2、絕緣試樣浸油中擊穿或耐壓試驗
2, insulation breakdown or compressive specimen immersed in test
3、絕緣試樣空氣中階梯擊穿或耐壓試驗
3, insulation test of the air in the air of the ladder breakdown or pressure test
4、絕緣試樣浸油中階梯擊穿或耐壓試驗
The 4 step, the insulation breakdown or compressive specimen immersed in test
五、試驗軟件:
Five, test software:
1、在試驗過程中可對升壓擊穿過程繪制實時曲線,每次試驗的升壓曲線都由不同顏色構(gòu)成,試驗結(jié)束后可疊加對比材料的試驗數(shù)據(jù)重復(fù)性。
1, in the process of the test can be used to draw real-time curve of the boost breakdown process, each test of the boost curve is composed of different colors, test data after the end of the experiment can be superimposed on the experimental data repeatability.
2、可以隨時調(diào)取當(dāng)前及歷史試驗數(shù)據(jù)進(jìn)行查看,編輯及修改參數(shù)。
2, can obtain the current and historical test data to view, edit and modify the parameters.
3、試驗過程中可以隨時修改試驗條件及存儲路徑及自動存儲試驗結(jié)果。
3. The test conditions and the storage path and automatic storage test results can be modified at any time in the process of testing.
4、試驗過程中,可隨時通過軟件決定本次試驗是否有效,方便篩選試驗結(jié)果。
4, the test process, the software can be determined at any time by the test is effective, easy to screen test results.
六、安全保護(hù)功能:
Six, safety protection function:
1、試驗在試驗箱中進(jìn)行,試驗箱門打開時電源加不到高壓變壓器輸入端,即高壓側(cè)無電壓。100KV測試設(shè)備高壓電極距離試驗箱壁的zui近距離大于270mm,50KV測試設(shè)備高壓電極距離試驗箱壁的zui近距離大于250mm,試驗時即使人接觸箱壁也不會有危險。
1, the test is carried out in the test chamber, and the power supply is not added to the input side of the high voltage transformer when the door is opened, that is, the high voltage side without the voltage. The nearest distance of the high voltage electrode distance of the test equipment is more than 270mm, and the distance of the 50KV test equipment is more than 100KV, the wall of the test chamber is more than 250mm.
2、介電強(qiáng)度測試設(shè)備要安裝單獨的保護(hù)地線。接保護(hù)地線,主要是減少試樣擊穿時對周圍產(chǎn)生的較強(qiáng)的電磁干擾。也可避免控制計算機(jī)失控。
2, the dielectric strength test equipment to install a separate protective ground wire. Earth wire, mainly to reduce the sample breakdown of the strong electromagnetic interference around the. Can also avoid control computer control.
3、電壓擊穿測試儀器的電路設(shè)有多項保護(hù)措施,主要有:過流保護(hù)、過壓保護(hù)、漏電保護(hù)、短路保護(hù)、直流試驗放電報警,放電裝置等。
3, voltage breakdown testing instrument circuit with a number of protective measures, mainly include: over current protection, over-voltage protection, leakage protection, short circuit protection, DC test discharge warning, discharge device, etc..
產(chǎn)品咨詢
歡迎您關(guān)注我們的微信公眾號了解更多信息
電話
微信掃一掃